Delayed cardiac perforation by defibrillator lead placed in the right ventricular outflow tract resulting in massive pericardial effusion.

Lau Ernest W, Shannon Heather J, McKavanagh Peter

Клинический случайСтатьяПеревод в очереди

Аннотация

A 76-year-old man received a dual-chamber implantable cardioverter defibrillator (ICD), with the defibrillator lead positioned within the right ventricular outflow tract. The lead parameters at the time of implantation were satisfactory and the post-procedure chest X-ray showed the leads were in place. The patient was cardioverted from atrial fibrillation during defibrillation threshold testing and commenced on anticoagulation immediately. One month post implantation, he experienced multiple ventricular tachycardia episodes all successfully treated with antitachycardia pacing and shocks by his ICD, but he fell and hit his chest against a hard surface during one of these attacks. He developed a massive pericardial effusion and computed tomography confirmed cardiac perforation by the defibrillator lead. Pericardiocentesis was performed and the defibrillator lead replaced with a different model positioned at the right ventricular apex. The patient made an uneventful recovery. The management and avoidance of delayed cardiac perforation by transvenous leads were discussed.

Для цитирования: Lau Ernest W, Shannon Heather J, McKavanagh Peter. Delayed cardiac perforation by defibrillator lead placed in the right ventricular outflow tract resulting in massive pericardial effusion.. Pacing and clinical electrophysiology : PACE. 2008. DOI: 10.1111/j.1540-8159.2008.01240.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.