[Treatment of proximal humerus fractures: relative position of different locking plates to the axillary nerve].
Аннотация на английскомСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ninck J, Heck S, Gick S, Koebke J, Pennig D, Dargel J
Аннотация
OBJECTIVES: Placement of a proximal humerus locking plate through a percutaneous transdeltoid approach bears the advantages of a minimally invasive approach but may compromise the anterior branches of the axillary nerve. This anatomic study aimed to develop a risk profile for 6 types of modern proximal humerus locking plates as to their interference with the axillary nerve.
MATERIALS AND METHODS: In this study six different implants (Arthrex®, DePuy®, Königsee®, Smith & Nephew®, Stryker® and Synthes®) were placed on the intact proximal humerus of 33 embalmed cadaveric upper extremities and the relative positioning between the axillary nerve and the screw holes was determined.
RESULTS: All locking plates displayed an area of risk which concerned 3 out of 7 (Arthrex®), 4 out of 10 (DePuy®), 2 out of 9 (Königsee®), 3 out of 11 (Smith & Nephew®), 3 out of 11 (Stryker®) and 6 out of 12 (Synthes®) screw holes of the plate.
CONCLUSIONS: Using the anterolateral percutaneous deltoid splitting approach the relative position of the axillary nerve to the holes of a specific implant is of relevance for avoidance of iatrogenic lesions to the nerve.
MATERIALS AND METHODS: In this study six different implants (Arthrex®, DePuy®, Königsee®, Smith & Nephew®, Stryker® and Synthes®) were placed on the intact proximal humerus of 33 embalmed cadaveric upper extremities and the relative positioning between the axillary nerve and the screw holes was determined.
RESULTS: All locking plates displayed an area of risk which concerned 3 out of 7 (Arthrex®), 4 out of 10 (DePuy®), 2 out of 9 (Königsee®), 3 out of 11 (Smith & Nephew®), 3 out of 11 (Stryker®) and 6 out of 12 (Synthes®) screw holes of the plate.
CONCLUSIONS: Using the anterolateral percutaneous deltoid splitting approach the relative position of the axillary nerve to the holes of a specific implant is of relevance for avoidance of iatrogenic lesions to the nerve.