MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/bot.0000000000002843

Медицинская библиотека · Journal of orthopaedic trauma · 2024

Safety of Magnetic Resonance Imaging in Orthopaedic Trauma Patients With External Fixation: A Two-Center Case Series.

Borgida Jacob S, Wagner Robert Kaspar, Wong Alice W, Yee Seonghwan, Husseini Jad, Aneja Arun, Harris Mitchel B, Ly Thuan V

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVES

To report on adverse events during magnetic resonance imaging (MRI) in patients with external fixators.

DESIGN

Retrospective case series.

SETTING

Two Level 1 trauma centers.

PATIENT SELECTION CRITERIA: Patients with external fixators on the appendicular skeleton or pelvis undergoing MRI between January 2005 and September 2023.

OUTCOME MEASURES AND COMPARISONS: Adverse events, defined as any undesirable event associated with the external fixator being inside or outside the MRI bore during imaging, including (subjective) heating, displacement or pullout of the external fixator, or early MRI termination for any reason.

RESULTS

A total of 97 patients with 110 external fixators underwent at least one MRI scan with an external fixator inside or outside of the MRI bore. The median age was 51 years (interquartile range: 39-63) and 56 (58%) were male. The most common external fixator locations were the ankle (24%), knee (21%), femur (21%), and pelvis (19%). The median duration of the MRI was 40 minutes (interquartile range: 26-58), 86% was performed using 1.5-Tesla MRI, and 14% was performed using 3.0-Tesla MRI. Ninety-five percent of MRI was performed for the cervical spine/head. Two MRI scans (1.6%), one of the shoulder and one of the head and cervical spine, with the external fixator outside of the bore were terminated early because of patient discomfort. There were no documented events of displacement or pullout of the external fixator.

CONCLUSIONS

These findings suggest that MRI scans of the (cervical) spine and head can be safely obtained in patients with external fixators on the appendicular skeleton or pelvis. Given the low numbers of MRI scans performed with the external fixator inside the bore, additional studies are necessitated to determine the safety of this procedure. The results from this study can aid orthopaedic surgeons, radiologists, and other stakeholders in developing local institutional guidelines on MRI scanning with external fixators in situ.

LEVEL OF EVIDENCE: Prognostic Level III. See Instructions for Authors for a complete description of levels of evidence.

Для цитирования: Borgida Jacob S, Wagner Robert Kaspar, Wong Alice W, Yee Seonghwan, Husseini Jad, Aneja Arun, Harris Mitchel B, Ly Thuan V. Safety of Magnetic Resonance Imaging in Orthopaedic Trauma Patients With External Fixation: A Two-Center Case Series.. Journal of orthopaedic trauma. 2024. DOI: 10.1097/bot.0000000000002843.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.