Minimal invasive coronary sinus lead reposition technique for the treatment of phrenic nerve stimulation.

Szilágyi Szabolcs, Merkely Béla, Zima Endre, Kutyifa Valentina, Szucs Gábor, Fülöp Gábor, Molnár Levente, Szabolcs Zoltán, Gellér László

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

AIMS: Phrenic nerve stimulation (PNS), which is often intolerable for the patient, is a known complication of resynchronization therapy. We describe a new, minimal invasive method for treating PNS.

METHODS AND

RESULTS

Untreatable PNS was found in nine cardiac resynchronization therapy patients with distal coronary sinus (CS) lead position 6 +/- 6 (0.5-17) months after the implantation. Ablation catheter and Amplatz Left 2 type guiding catheter were introduced into the right atrium via the right femoral vein. Coronary sinus was cannulated with the Amplatz catheter, and on a normal guide wire, a coronary stent was introduced beside the lead into the side branch in seven cases or a bigger stent into the CS in two patients. The ablation catheter was looped around the CS lead in the atrium with bent tip and was drawn backward together with the CS electrode. New lead positions were evaluated with electrophysiological measurements, and the suitable position was stabilized with inflation of the stent. Pericardial effusion was not detected on post-operative echocardiography. After repositioning, suitable pacing parameters were registered (threshold: 1.6 +/- 1.1 V; 0.5 ms, impedance: 565 +/- 62 ohm). Phrenic nerve stimulation was not found with 7.5 V; 1.5 ms pacing. During follow-up (7.7 +/- 4.6 months), stable pacing threshold and impedance values were measured; transient and reprogrammable PNS was present in only one patient.

CONCLUSION

Coronary sinus electrode reposition using the femoral approach seems to be a safe and effective procedure, which means smaller burden for the patients compared with the established reposition operation. The technique can be used successfully if the CS lead is in a distal position.

Для цитирования: Szilágyi Szabolcs, Merkely Béla, Zima Endre, Kutyifa Valentina, Szucs Gábor, Fülöp Gábor, Molnár Levente, Szabolcs Zoltán, Gellér László. Minimal invasive coronary sinus lead reposition technique for the treatment of phrenic nerve stimulation.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2008. DOI: 10.1093/europace/eun207.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.