DOI 10.1080/09553002.2025.2494554
LBDNet inter-laboratory comparison at high doses of ionizing radiation using the dicentric plus caffeine assay.
Аннотация
PURPOSE: To assess the performance of the LBDNet laboratories in estimating dose over 5 Gy of ionizing radiation using the dicentric chromosome plus caffeine assay.
MATERIALS AND
METHODS
Dose-response curve fitting: Peripheral blood was irradiated in vitro between 5 and 25 Gy. Then, the DC plus caffeine assay was carried out. Thirteen laboratories received and analyzed metaphase images. The linear dose-response curve was fitted for each laboratory. Dose estimation was performed analyzing coded metaphase images from three different irradiated samples (7.5, 15, 20 Gy) and using the fitted curve from every laboratory.
RESULTS
The dose estimation accuracy was within the expected dose ranges. The 76.9%, 84.6% and 69.2% of the estimated doses fell into the ± 20% of the true radiation dose. The 92.3%, 92.3%, and 61.5% of the 95% of the confidence interval of the estimated doses included the true radiation dose. The trueness was 0.9%, 4.4% and 9.6%. The Coefficients of Variation of the estimated doses were 14.5%, 16.1% and 17.8%. Results from only one laboratory were deemed questionable for dose estimation, based on the Z-score derived from robust methods.
CONCLUSION
The intercomparison study yielded satisfactory results; however, dose estimation accuracy tended to decrease, and variability between laboratory results increased as the dose level rose.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.