Thrombosis research · 2026

Pre-analytical storage and reagent-dependent sensitivity as sources of variability in prothrombin time and activated partial thromboplastin time measurements.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

van Dam A D, Bioch M M, Jeninga E H, de Kok J B, Rijpma S R, Thelen M H M, Henskens Y M C, van Oerle R G M, Stroobants A K, Coagulation Section of the Dutch Foundation for Quality Assessment in Medical Laboratory Diagnostics (SKML)

Аннотация

BACKGROUND: Prothrombin time (PT) and activated partial thromboplastin time (aPTT) are widely used coagulation assays whose results can vary due to pre-analytical and analytical factors. Understanding these sources of variability is essential for reliable laboratory practice and external quality assessment (EQA).

OBJECTIVES: This study evaluates the impact of pre-analytical storage and method-dependent sensitivity on PT and aPTT measurements and provides guidance for EQA programmes and clinical interpretation.

METHODS: Plasma samples from 42 healthy volunteers were analysed within 4 h of collection or after storage at -20 °C, -80 °C, or snap-frozen in liquid nitrogen (LN₂) and subsequent storage at -80 °C. Additionally, pooled plasma was spiked with direct oral anticoagulants, heparins, and lupus anticoagulant antibodies to assess reagent-specific sensitivity. PT and aPTT were measured on multiple routine platforms using commonly used reagents from 4 different manufacturers.

RESULTS: Storage at -80 °C or LN₂ preserved PT and aPTT values, whereas storage at -20 °C caused variable prolongation. In healthy volunteers, PT and aPTT showed marked inter-method variability. Sensitivity to anticoagulants and lupus anticoagulant also differed considerably between reagents, with the largest variation observed for edoxaban. This study provides an overview of inter-method variability, which may affect clinical decision-making and inter-laboratory comparability.

CONCLUSIONS: PT and aPTT are screening tests whose accurate interpretation depends on optimal sample handling and awareness of reagent- and method-specific sensitivities. These differences can affect clinical decision-making in anticoagulated patients. EQA programmes could increase their value by using frozen plasma, including clinically relevant anticoagulants, and implementing case-based scenarios rather than reporting only clotting times.