Injury to the axillary artery, a complication of fixation using a locking plate.
Клинический случайСтатьяПеревод в очереди
Аннотация
The proximal humerus internal locking system is an internal fixation device consisting of a low-profile plate and locking screws, which is used for the fixation of three- and four-part fractures of the proximal humerus. We describe a case in which the screws cut out of the humeral head causing injury to the axillary artery necessitating urgent removal of the implant.
Для цитирования: Khunda A, Stirrat A N, Dunlop P. Injury to the axillary artery, a complication of fixation using a locking plate.. The Journal of bone and joint surgery. British volume. 2007. DOI: 10.1302/0301-620x.89b11.19264.