Single-center comparison of gel microcolumn and solid-phase methods for antibody screening.
Аннотация
Our facility changed antibody screening methods from a gel microcolumn-based test (ID-Micro Typing System Gel TEst; Ortho Clinical Diagnostics, Inc., Raritan, NJ) to an automated solid-phase test (Galileo/Capture-R Ready Screen [I and II], Immucor, Inc., Norcross, GA). To determine whether detection rates for commonly encountered clinically significant red blood cell antibodies differed as a consequence of this change, preimplementation and postimplementation antibody identification records were retrospectively reviewed. A statistically significant difference in the percentage of positive screening tests during the gel microcolumn testing period (73,903 total screens, 1.56% confirmed positive) versus the solid0-phase screening period (80,242 total screens, 1.81% confirmed positive; p< 0.0002) was observed . The number of antibodies to K identified was significantly lower with solid phase that with gel (27% decrease; p=0.004). It is unknown whether there is a statistical difference in delayed or hemolytic transfusion reaction rates as this was not evaluated.