MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0358184

Медицинская библиотека · PloS one · 2026

Impact of dual-layer spectral CT-derived stopping power ratio on single-energy CT-based carbon-ion treatment planning for thoracic cancer.

Yoon Euntaek, Lee Sung Hyun, Jin Hyeongmin, Jang Tae Ho, Yoon Sungjin, Choun Hyung Jin, Yoon Soon Ho, Goo Jin Mo, Park Jong Min

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the impact of stopping-power ratio (SPR) predictions derived from dual-layer detector-based spectral CT (DLCT) on conventional single-energy CT (SECT)-based carbon-ion treatment planning for thoracic cancer.

METHODS

Twenty patients with thoracic cancer who underwent non-contrast DLCT were retrospectively included. DLCT-based SPR images were generated from effective atomic number and relative electron density maps using the Bethe equation. SECT-based SPR was derived using a Hounsfield unit look-up table generated via stoichiometric calibration of a commercial electron density phantom. Using the same phantom, a phantom-based prediction assessment was performed by comparing SECT- and DLCT-based SPR predictions with theoretical SPR values for the inserts. For patient analysis, normal-tissue SPR values derived from SECT and DLCT were compared. Carbon-ion treatment plans were first generated based on SECT images and then recalculated using DLCT-derived SPR images. Range differences in the beam's-eye view and dose-volume histogram parameters for the planning target volume and organs at risk were evaluated.

RESULTS

In the phantom-based assessment, the root-mean-square error relative to the theoretical SPR was 5.57% for SECT and 2.83% for DLCT. In patient images, mean SPR differences between SECT and DLCT were 3.62% for the lungs, 1.93% for the esophagus, 0.04% for the heart, 0.93% for the spinal cord, and 1.53% for the ribs. Across the patient cohort, the mean recalculated range difference was 0.71 ± 0.89 mm, corresponding to a relative difference of 0.60 ± 0.67%. No statistically significant differences were observed in target dose metrics, whereas selected organ-at-risk metrics showed statistically significant differences between SECT-based plans and DLCT-based recalculations.

CONCLUSION

DLCT-derived SPR prediction showed better agreement with theoretical phantom SPR values in the phantom-based prediction assessment and produced tissue-dependent SPR differences in thoracic patient images compared with conventional SECT-based prediction. Recalculation on DLCT-derived SPR maps resulted in modest but systematic range differences and limited changes in selected dose metrics. These results may inform the implementation and future independent validation of patient-specific DLCT-based SPR workflows in carbon-ion treatment planning for thoracic cancer.

Для цитирования: Yoon Euntaek, Lee Sung Hyun, Jin Hyeongmin, Jang Tae Ho, Yoon Sungjin, Choun Hyung Jin, Yoon Soon Ho, Goo Jin Mo, Park Jong Min. Impact of dual-layer spectral CT-derived stopping power ratio on single-energy CT-based carbon-ion treatment planning for thoracic cancer.. PloS one. 2026. DOI: 10.1371/journal.pone.0358184.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.