OBJECTIVES

The objective of this study was to evaluate the reliability and accuracy of the combined use of MALDI-TOF MS bacterial identification and the Vitek-2 Compact antimicrobial susceptibility testing (AST) directly from positive blood cultures.

METHODS

Direct identification by MALDI-TOF MS and AST were performed in parallel to the standard methods in all positively flagged blood cultures bottles during the study period.

RESULTS

Three hundred and twenty four monomicrobial positive blood cultures were included in the present study, with 257 Gram-negative and 67 Gram-positive isolates. MALDI-TOF MS identification directly from blood bottles reported the correct identification for Enterobacteriaceae in 97.7%, non-fermentative Gram-negative bacilli 75.0%, Staphylococcus aureus 75.8%, coagulase negative staphylococci 63.3% and enterococci 63.3%. A total 6156 isolate/antimicrobial agent combinations were tested. Enterobacteriaceae group and non-fermentative Gram-negative Bacilli showed an agreement of 96.67% and 92.30%, respectively, for the Gram-positive cocci the overall agreement found was 97.84%.

CONCLUSIONS

We conclude that direct identification by MALDI-TOF and inoculation of Vitek-2 Compact AST with positive blood culture bottles yielded very good results and decreased time between initial inoculation of blood culture media and determination of the antibiotic susceptibility for Gram-negative rods and Gram-positive cocci causing bacteremia.

Для цитирования: Romero-Gómez María-Pilar, Gómez-Gil Rosa, Paño-Pardo Jose Ramón, Mingorance Jesús. Identification and susceptibility testing of microorganism by direct inoculation from positive blood culture bottles by combining MALDI-TOF and Vitek-2 Compact is rapid and effective.. The Journal of infection. 2012. DOI: 10.1016/j.jinf.2012.08.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.