Stability of One-Step Spray-on Splint for Lower Extremity Fractures During Splinting, MEDEVAC, and Impact.

Hobayan C Grace P, Bates Nathaniel A, Heyniger John, Alzouhayli Kenan, Piscitani Franco, Haider Clifton R, Felton Christopher, Groth Adam T, Martin Kevin D

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Military transport can induce whole-body vibrations, and combat almost always involves high impact between lower extremities and the ground. Therefore, robust splinting technology is necessary for lower extremity fractures in these settings. Our team compared a novel one-step spray-on foam splint (FastCast) to the current military standard structured aluminum malleable (SAM) splint.

MATERIALS AND

METHODS

Ten cadaveric specimens were subjected to complete tibia/fibula osteotomy. Specimens were fitted with custom accelerometer and gyroscope sensors superior and inferior to the fracture line. Each specimen underwent fracture and splinting from a standard of care SAM splint and an experimental FastCast spray foam splint in a randomized order. Each specimen was manually transported to an ambulance and then released from a 1 meter height to simulate impact. The custom sensors recorded accelerations and rotations throughout each event. Repeated-measures Friedman tests were used to assess differences between splint method within each event and between sensors within each splint method.

RESULTS

During splinting, overall summation of change and difference of change between sensors for accelerations and rotations were greater for SAM splints than FastCast across all axes (P ≤ 0.03). During transport, the range of acceleration along the linear superior/inferior axis was greater for SAM splint than FastCast (P = 0.02), as was the range of rotation along the transverse plane (P < 0.01). On impact, the summation of change observed was greater for SAM splint than FastCast with respect to acceleration and rotation on the posterior/anterior and superior/inferior axes (P ≤ 0.03), and the cumulative difference between superior and inferior sensors was greater for SAM than FastCast with respect to anterior-axis rotation (P < 0.05).

CONCLUSION

FastCast maintains stabilization of fractured lower extremities during transport and impacts to a significantly greater extent than SAM splints. Therefore, FastCast can potentially reduce the risk of fracture complications following physical stressors associated with combat and extraction.

Для цитирования: Hobayan C Grace P, Bates Nathaniel A, Heyniger John, Alzouhayli Kenan, Piscitani Franco, Haider Clifton R, Felton Christopher, Groth Adam T, Martin Kevin D. Stability of One-Step Spray-on Splint for Lower Extremity Fractures During Splinting, MEDEVAC, and Impact.. Military medicine. 2024. DOI: 10.1093/milmed/usae001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.