Optimum design parameters for a therapist-constructed positive-expiratory-pressure therapy bottle device.

Mestriner Régis Gemerasca, Fernandes Rafael Oliveira, Steffen Luís Carlos, Donadio Márcio Vinícius Fagundes

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor.

METHODS

We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm).

RESULTS

With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing > or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure.

CONCLUSIONS

To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be > or = 8 mm inner diameter, and the air-escape orifice must be > or = 8 mm.

Для цитирования: Mestriner Régis Gemerasca, Fernandes Rafael Oliveira, Steffen Luís Carlos, Donadio Márcio Vinícius Fagundes. Optimum design parameters for a therapist-constructed positive-expiratory-pressure therapy bottle device.. Respiratory care. 2009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.