MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jpeds.2026.115247

Медицинская библиотека · The Journal of pediatrics · 2026

Non-Invasive Optical Quantification of Cerebral Tissue Oxygenation in Children with Single-Ventricle Congenital Heart Disease.

Zhao Hongting, Davis Caitlyn, Anderson Darci, Bayle Briston, Schulz Alexis, Ranieri Nicolina, Padiyath Asif, Mirreh Manal, Matthews Lea, Majeski Joseph B, Forti Rodrigo Menezes, Ko Tiffany S, Yodh Arjun G, White Brian R, Baker Wesley B, Lynch Jennifer M

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare the accuracy of frequency-domain near-infrared spectroscopy (FD-NIRS) with clinically available optical measures of cerebral tissue oxygenation (StO2) and arterial saturation (SaO2).

STUDY

DESIGN

We enrolled 24 children with single-ventricle congenital heart defects scheduled for cardiac catheterization. Intraoperative FD-NIRS and continuous-wave near-infrared spectroscopy (CW-NIRS) measurements of StO2 were bilaterally obtained on the forehead. SaO2 was simultaneously acquired with FD-NIRS and pulse oximetry. Noninvasive StO2 and SaO2 were compared with invasive superior vena cava (SVC) and descending aorta saturation measurements. These measurements were also used to compute cerebral tissue oxygen extraction fraction (OEF).

RESULTS

The FD-NIRS StO2 was strongly correlated with invasive venous oxygen saturation (SvO2) (R = 0.74, P < .001; bias, -2.8%). However, correlation between CW-NIRS StO2 and invasive SvO2 was weaker and not significant (R = 0.29, P = .34; bias, -2.5%). Invasive SaO2 was correlated with FD-NIRS (R = 0.84, P < .001; bias, -3.2%) and pulse oximetry (R = 0.68, P = .001; bias, 0.2%) metrics of SaO2, but pulse oximetry showed larger error at low saturations. Finally, FD-NIRS OEF and invasive OEF were strongly correlated (R = 0.61, P = .01; bias, 0.09).

CONCLUSIONS

Our study suggests that FD-NIRS provides more accurate quantification of StO2 and SaO2 than CW-NIRS and pulse oximetry in patients with single-ventricle heart defects. We also demonstrated and validated stand-alone OEF quantification using FD-NIRS.

Для цитирования: Zhao Hongting, Davis Caitlyn, Anderson Darci, Bayle Briston, Schulz Alexis, Ranieri Nicolina, Padiyath Asif, Mirreh Manal, Matthews Lea, Majeski Joseph B, Forti Rodrigo Menezes, Ko Tiffany S, Yodh Arjun G, White Brian R, Baker Wesley B, Lynch Jennifer M. Non-Invasive Optical Quantification of Cerebral Tissue Oxygenation in Children with Single-Ventricle Congenital Heart Disease.. The Journal of pediatrics. 2026. DOI: 10.1016/j.jpeds.2026.115247.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.