Management of a Broken PRECICE Femoral Nail at an Ununited Distraction Osteogenesis Site: A Case Report.

Hidden Krystin A, Dahl Mark T, Ly Thuan V

Клинический случайСтатьяПеревод в очереди

Аннотация

CASE: A 20-year-old man with a history of right lower extremity fibular hemimelia previously treated with PRECICE femoral nail lengthening presented with a broken magnetic nail and a displaced fracture through an ununited distraction osteogenesis site. Using a combination of techniques, we removed the broken implant while maintaining the achieved limb length and preserving the native biology without bone grafting.

CONCLUSION

The unique challenges associated with the removal of a broken PRECICE femoral nail are described, with a technique for implant removal that preserves the achieved length, the innate biology of the distraction osteogenesis site, and promoting union without bone grafting.