Tremor Artifact Presenting as Pseudo-Ventricular Tachycardia.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Haiy Ata Ul, Basit Muhammad Abdul, Mirza Hasan Mahmood, Wasim Aaima, Jalil Ahmad, Bajwa Ata
Аннотация
BACKGROUND: Electrocardiographic artifacts can present as pseudo-arrhythmia, such as ventricular tachycardia (VT), leading to unnecessary interventions.
CASE SUMMARY: A 47-year-old woman with hypertension, asthma, obstructive sleep apnea, lupus/Sjögren disease, and schizoaffective disorder was admitted for failure to thrive. Telemetry suggested polymorphic and monomorphic VT, prompting a rapid response. She was asymptomatic and hemodynamically stable, with a pulse 93/min. A baseline tremor was noted. Post-rapid-response electrocardiogram revealed an normal sinus rhythm with a QTc interval of 460 milliseconds. Preserved QRS complexes were seen to be embedded within the observed VT on telemetry review. Identification of the artifact pattern prevented further unwarranted interventions.
DISCUSSION: Tremor-induced artifacts can present as compelling VT mimics. Stable hemodynamics, concealed QRS complexes, and bedside examination are crucial in distinguishing pseudo-arrhythmia from true arrhythmia.
TAKE-HOME MESSAGES: Electrocardiographic artifacts should be considered in clinically stable patients with telemetry VT alarms. Careful holistic approach to such alarms can avoid unnecessary invasive procedures and antiarrhythmic use.
CASE SUMMARY: A 47-year-old woman with hypertension, asthma, obstructive sleep apnea, lupus/Sjögren disease, and schizoaffective disorder was admitted for failure to thrive. Telemetry suggested polymorphic and monomorphic VT, prompting a rapid response. She was asymptomatic and hemodynamically stable, with a pulse 93/min. A baseline tremor was noted. Post-rapid-response electrocardiogram revealed an normal sinus rhythm with a QTc interval of 460 milliseconds. Preserved QRS complexes were seen to be embedded within the observed VT on telemetry review. Identification of the artifact pattern prevented further unwarranted interventions.
DISCUSSION: Tremor-induced artifacts can present as compelling VT mimics. Stable hemodynamics, concealed QRS complexes, and bedside examination are crucial in distinguishing pseudo-arrhythmia from true arrhythmia.
TAKE-HOME MESSAGES: Electrocardiographic artifacts should be considered in clinically stable patients with telemetry VT alarms. Careful holistic approach to such alarms can avoid unnecessary invasive procedures and antiarrhythmic use.