MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/ons/opz003

Медицинская библиотека · Operative neurosurgery (Hagerstown, Md.) · 2019

First-in-Human Experience With Integration of a Hydrocephalus Shunt Device Within a Customized Cranial Implant.

Gordon Chad R, Wolff Amir, Santiago Gabriel F, Liebman Kenneth, Veznedaroglu Erol, Vrionis Frank D, Huang Judy, Brem Henry, Luciano Mark

Клинический случайСтатьяVideo-Audio MediaПеревод в очереди

Аннотация

BACKGROUND

Implantable shunt devices are critical and life saving for hydrocephalus patients. However, these devices are fraught with high complication rates including scalp dehiscence, exposure, and extrusion. In fact, high shunt valve profiles are correlated with increased complications compared to those with lower profiles. As such, we sought a new method for integrating shunt valves for those challenging patients presenting with scalp-related complications.

OBJECTIVE

To safely implant and integrate a hydrocephalus shunt valve device within a customized cranial implant, in an effort to limit its high-profile nature as a main contributor to shunt failure and scalp breakdown, and at the same time, improve patient satisfaction by preventing contour deformity.

METHODS

A 64-yr-old male presented with an extruding hydrocephalus shunt valve and chronic, open scalp wound. The shunt valve was removed and temporary shunt externalization was performed. He received 2 wk of culture-directed antibiotics. Next, a contralateral craniectomy was performed allowing a new shunt valve system to be implanted within a low-profile, customized cranial implant. All efforts were made, at the patient's request, to decrease the high-profile nature of the shunt valve contributing to his most recent complication.

RESULTS

First-in-human implantation was performed without complication. Postoperative shunt identification and programming was uncomplicated. The high-profile nature of the shunt valve was decreased by 87%. At 10 mo, the patient has experienced no complications and is extremely satisfied with his appearance.

CONCLUSION

This first-in-human experience suggests that a high-profile hydrocephalus shunt device may be safely integrated within a customized cranial implant.

Для цитирования: Gordon Chad R, Wolff Amir, Santiago Gabriel F, Liebman Kenneth, Veznedaroglu Erol, Vrionis Frank D, Huang Judy, Brem Henry, Luciano Mark. First-in-Human Experience With Integration of a Hydrocephalus Shunt Device Within a Customized Cranial Implant.. Operative neurosurgery (Hagerstown, Md.). 2019. DOI: 10.1093/ons/opz003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.