DOI 10.1007/s11604-026-01964-0

Медицинская библиотека · Japanese journal of radiology · 2026

High-resolution chest CT using 1024-matrix reconstruction: phantom and clinical evaluation of image quality and post-processing capability.

Hayashi Kana, Ozawa Yoshiyuki, Ohkubo Hirotsugu, Okuda Katsuhiro, Kunitomo Hiroshi, Watanabe Seita, Kato Katsuya, Hiwatashi Akio

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the image quality and clinical utility of high-resolution computed tomography (CT) images reconstructed using post-processing with a 1024 × 1024 matrix.

MATERIALS AND

METHODS

Four CT image types were reconstructed: type A (512 × 512 matrix, 1-mm slice thickness), type B (1024 × 1024 matrix, 1-mm slice thickness), type C (1024 × 1024 matrix, 0.5-mm slice thickness), and type D (retargeted zoomed images; 512 × 512 matrix, 1-mm slice thickness). Type D images served as the reference standard with a field of view (FOV) half that of the other image types. Phantom image quality was objectively assessed using the task transfer function (TTF), noise power spectrum (NPS), and attenuation profile analysis. In the clinical study, two radiologists independently evaluated image quality in 93 patients, including abnormal lung findings, normal structures, and subjective image noise.

RESULTS

In the phantom study, post-processing 1024 × 1024 matrix reconstruction derived from the same sinogram as conventional 512 × 512 reconstruction improved spatial resolution while maintaining comparable noise characteristics at the same slice thickness. Clinically, image quality of normal structures and abnormal lesions was significantly higher in the order of type C > type B > type A (p < 0.0001). Subjective image noise followed the same order. Image quality and noise levels of type B images were comparable to those of type D images.

CONCLUSION

Post-processing 1024 × 1024 matrix reconstruction improves spatial resolution and provides clinically useful high-resolution CT images comparable to retargeted zoomed reconstruction without the limitation of a reduced FOV.

SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s11604-026-01964-0.

Для цитирования: Hayashi Kana, Ozawa Yoshiyuki, Ohkubo Hirotsugu, Okuda Katsuhiro, Kunitomo Hiroshi, Watanabe Seita, Kato Katsuya, Hiwatashi Akio. High-resolution chest CT using 1024-matrix reconstruction: phantom and clinical evaluation of image quality and post-processing capability.. Japanese journal of radiology. 2026. DOI: 10.1007/s11604-026-01964-0.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.