Effectiveness of a flow confirmation study of a central venous port of the upper arm versus the chest wall in patients with suspected system-related mechanical complications.

Tokue Hiroyuki, Tokue Azusa, Tsushima Yoshito

СтатьяПеревод в очереди

Аннотация

BACKGROUND

If mechanical complications associated with a central venous port (CVP) system are suspected, evaluation with a flow confirmation study (FCS) using fluorescence fluoroscopy or digital subtraction angiography should be performed. Evaluations of mechanical complications related to CVP of the chest wall using FCS performed via the subclavian vein have been reported. However, the delayed complications of a CVP placed in the upper arm have not been sufficiently evaluated in a large population. We evaluated the effectiveness of FCS of CVPs implanted following percutaneous cannulation of the subclavian (chest wall group) or brachial (upper arm group) vein.

METHODS

A CVP was implanted in patients with advanced cancer requiring chemotherapy. FCS was performed if there were complaints suggestive of CVP dysfunction when initiating chemotherapy.

RESULTS

CVPs were placed in the brachial vein in 390 patients and in the subclavian vein in 800 patients. FCS was performed in 26/390 (6.7%) patients in the upper arm group and 40/800 (5.0%) patients in the chest wall group. The clinical characteristics of the patients were similar in both groups. The duration of CVP implantation until FCS was significantly shorter in the upper arm group (136 ± 96.6 vs. 284 ± 260, p = 0.022). After FCS, the incidence of CVP removal/reimplantation being deemed unnecessary was higher in the upper arm group (21/26 [80.8%] vs. 26/40 [65.0%], p = 0.27). In the upper arm group, no cases of catheter kinking or catheter-related injury were observed, and the incidence of temporary obstruction because of blood clots that could be continued using CVP was significantly higher than that in the chest wall group (10/26 [38.5%] vs. 4/40 [10.0%], p = 0.012).

CONCLUSIONS

FCS was effective in evaluating CVP system-related mechanical complications and deciding whether removal and reimplantation were required in both groups.

Для цитирования: Tokue Hiroyuki, Tokue Azusa, Tsushima Yoshito. Effectiveness of a flow confirmation study of a central venous port of the upper arm versus the chest wall in patients with suspected system-related mechanical complications.. World journal of surgical oncology. 2022. DOI: 10.1186/s12957-022-02565-7.