MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.arth.2016.06.053

Медицинская библиотека · The Journal of arthroplasty · 2016

Does Taper Size Have an Effect on Taper Damage in Retrieved Metal-on-Polyethylene Total Hip Devices?

Higgs Genymphas B, MacDonald Daniel W, Gilbert Jeremy L, Rimnac Clare M, Kurtz Steven M, Implant Research Center Writing Committee

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Taper design has been identified as a possible contributor to fretting corrosion damage at modular connections in total hip arthroplasty systems, but variations in as-manufactured taper interfaces may confound this analysis. This study characterized taper damage in retrievals with 2 different taper sizes but comparable taper surface finishes and investigated if fretting and corrosion damage is related to taper size in the context of a multivariable analysis for metal-on-polyethylene bearings.

METHODS

A total of 252 cobalt chromium femoral heads were identified in a collection of retrievals: 77 with taper A and 175 with taper B. Implantation time averaged 5.4 ± 6.0 years (range, 0-26 years). Explants were cleaned and analyzed using a 4-point semiquantitative method. Clinical and device factors related to head taper fretting corrosion damage were assessed using ordinal logistic regression with forward stepwise control. Components were then selected to create 2 balanced cohorts, matched on significant variables from the multivariable analysis.

RESULTS

Increased head offset (P < .001), longer implantation time (P = .002), heavier patients (P < .001), and more flexible tapers (P < .001) were associated with increased taper fretting and corrosion damage. When damage scores were compared between the balanced groups, no significant differences were found.

CONCLUSION

These results suggest that fretting and corrosion damage is insensitive to differences in taper size. The final model derived explains almost half of the fretting corrosion damage observed and identifies contributing factors that are consistent with other in vitro and retrieval studies.

Для цитирования: Higgs Genymphas B, MacDonald Daniel W, Gilbert Jeremy L, Rimnac Clare M, Kurtz Steven M, Implant Research Center Writing Committee. Does Taper Size Have an Effect on Taper Damage in Retrieved Metal-on-Polyethylene Total Hip Devices?. The Journal of arthroplasty. 2016. DOI: 10.1016/j.arth.2016.06.053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.