Inclinometer Reliability for Shoulder Ranges of Motion in Individuals With Subacromial Impingement Syndrome.

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Tozzo Marcela C, Ansanello Walter, Martins Jaqueline, Zatiti Salomão C A, de Oliveira Anamaria S

Аннотация

OBJECTIVE: This study aimed to determine the intra- and interrater reliability of active and passive range of motion in the shoulders of individuals with subacromial impingement syndrome using a digital inclinometer.

METHODS: The raters evaluated active and passive range of motion in the shoulder of 50 individuals with unilateral subacromial impingement syndrome in movements including flexion, abduction, extension, external rotation in a neutral position, external rotation with the arm at 90° of abduction, and internal rotation with the arm at 90° of abduction. The tests were performed by 2 examiners on the same day, with a 10-minute interval, and were repeated by 1 examiner after a 2- to 4-day interval. Reliability was analyzed using the intraclass correlation coefficient (ICC2,3).

RESULTS: There was moderate to excellent interrater (ICC2,3 = 0.50-0.95) and intrarater (ICC2,3 = 0.74-0.94) reliability. In the interrater analysis, the standard error of measurement (SEM) ranged from 4.1° to 10°, the percentage SEM (%SEM) ranged from 2% to 17%, and the minimum detectable change ranged from 9.5° to 23.4°. In the intrarater analysis, the SEM ranged from 4° to 9.2°, %SEM ranged from 3% to 14%, and the minimum detectable change ranged from 9.3° to 21.4°.

CONCLUSION: The digital inclinometer showed moderate to excellent reliability for measuring active and passive range of motion in shoulders with unilateral subacromial impingement syndrome.