DOI 10.1016/j.healun.2023.01.011
Improved clinical outcomes associated with the Impella 5.5 compared to the Impella 5.0 in contemporary cardiogenic shock and heart failure patients.
Аннотация
A redesigned surgically implanted heart pump incorporates several design changes from the prior device generation, but no published comparative data demonstrate if these changes translate to improved outcomes. We retrospectively compared clinical characteristics and outcomes, drawn from an FDA-mandated QA database, for contemporary patients treated with the Impella 5.5 or Impella 5.0 for acute myocardial infarction complicated by cardiogenic shock (AMICS), cardiomyopathy, or postcardiotomy cardiogenic shock (PCCS). A total of 1238 patients at 290 US sites were included for analysis. Patients receiving the Impella 5.5 had significantly higher survival through explant (i.e., successfully weaned or bridged to heart replacement therapy) than those receiving the Impella 5.0 in all 3 settings: AMICS (70.5% vs 56.8%; p = 0.005), cardiomyopathy (88.1% vs 76.9%; p = 0.001), and PCCS (76.1% vs 55.7%; p = 0.003). Duration of support was significantly longer for Impella 5.5 patients with AMICS (9.2 vs 6.1 days; p = 0.008) and cardiomyopathy (10.7 vs 8.1 days; p < 0.001).