A trial comparing continuous positive airway pressure (CPAP) devices in preterm infants.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Backes Carl H, Cooper Jennifer N, Notestine Jennifer L, Alfred Crystal M, Ball Molly K, Rivera Brian K, Lamp Jane M, Marzec Laura, Stenger Michael R, Moallem Mohannad, Miller Randy R, Naik Apurwa, Beer Lindsey J, Howard Christopher R, Welty Stephen E, Peter Richardson C, Hillman Noah H, Zupancic John A F, Stanberry Larissa I, Hansen Thomas N, Smith Charles V
Аннотация
OBJECTIVE: To test the hypothesis that infants born <30 weeks' gestation supported by Seattle-PAP will have lower rates of continuous positive airway pressure (CPAP) failure than infants supported with conventional, Fisher&Paykel-CPAP (FP-CPAP).
STUDY DESIGN: Randomized trial (3/2017-01/2019) at 5 NICUs. The primary outcome was CPAP failure; subgroup analyses (gestational age, receipt antenatal corticosteroids) were performed.
RESULTS: A total of 232 infants were randomized. Infants in the Seattle-PAP and FP-CPAP groups had mean gestational ages of 27.0 and 27.2 weeks, respectively. We observed no differences in rates of treatment failure between Seattle-PAP (40/112, 35.7%) and FP-CPAP (38/120, 31.7%; risk difference, 4.1%; 95% CI, -8.1-16.2; P = 0.51). Subgroup analysis indicated no differences in rates of CPAP failure. We observed no differences between the two groups in frequencies of adverse events or duration of respiratory support.
CONCLUSIONS: Among infants born <30 weeks' gestation, rates of CPAP failure did not differ between Seattle-PAP and FP-CPAP.
STUDY DESIGN: Randomized trial (3/2017-01/2019) at 5 NICUs. The primary outcome was CPAP failure; subgroup analyses (gestational age, receipt antenatal corticosteroids) were performed.
RESULTS: A total of 232 infants were randomized. Infants in the Seattle-PAP and FP-CPAP groups had mean gestational ages of 27.0 and 27.2 weeks, respectively. We observed no differences in rates of treatment failure between Seattle-PAP (40/112, 35.7%) and FP-CPAP (38/120, 31.7%; risk difference, 4.1%; 95% CI, -8.1-16.2; P = 0.51). Subgroup analysis indicated no differences in rates of CPAP failure. We observed no differences between the two groups in frequencies of adverse events or duration of respiratory support.
CONCLUSIONS: Among infants born <30 weeks' gestation, rates of CPAP failure did not differ between Seattle-PAP and FP-CPAP.