A trial comparing continuous positive airway pressure (CPAP) devices in preterm infants.
Аннотация
OBJECTIVE
To test the hypothesis that infants born <30 weeks' gestation supported by Seattle-PAP will have lower rates of continuous positive airway pressure (CPAP) failure than infants supported with conventional, Fisher&Paykel-CPAP (FP-CPAP).
STUDY
DESIGN
Randomized trial (3/2017-01/2019) at 5 NICUs. The primary outcome was CPAP failure; subgroup analyses (gestational age, receipt antenatal corticosteroids) were performed.
RESULTS
A total of 232 infants were randomized. Infants in the Seattle-PAP and FP-CPAP groups had mean gestational ages of 27.0 and 27.2 weeks, respectively. We observed no differences in rates of treatment failure between Seattle-PAP (40/112, 35.7%) and FP-CPAP (38/120, 31.7%; risk difference, 4.1%; 95% CI, -8.1-16.2; P = 0.51). Subgroup analysis indicated no differences in rates of CPAP failure. We observed no differences between the two groups in frequencies of adverse events or duration of respiratory support.
CONCLUSIONS
Among infants born <30 weeks' gestation, rates of CPAP failure did not differ between Seattle-PAP and FP-CPAP.