Bond strength of composite to alloy treated with bonding systems.
Аннотация
PURPOSE: The in-vitro bond strengths of a composite bonded to a nickel-chromium-beryllium alloy treated by eight bonding systems were measured after three storage conditions.
MATERIALS AND
METHODS
Nickel-chromium-beryllium alloy samples were treated by eight commercial bonding systems including adhesive composite cements, all-purpose bonding agents, and silica-coating systems. A composite was bonded to the alloy samples. Sample groups were stored in water for 24 hours at 23 degrees C, or thermocycled, or stored for 6 months at 23 degrees C and then debonded in tension.
RESULTS
Bond strengths after 24 hours storage were: adhesive composite cements, 14.2 to 22.1 MPa; all-purpose bonding agents, 11.4 to 14.6 MPa; and silica-coating systems, 18.6 to 20.2 MPa. Bond strengths after thermocycling were: adhesive composite cements, 12.6 to 20.6 MPa; all-purpose bonding agents, 9.9 to 17.7 MPa; and silica-coating systems, 11.1 to 19.2 MPa. Bond strengths after 6 months were: adhesive composite cements, 12.0 to 13.1 MPa; all-purpose bonding agents, 8.7 to 14.1 MPa; and silica-coating systems, 14.8 to 18.4 MPa.
CONCLUSIONS
Only two bonding systems showed decreased bond strength after thermocycling, as compared with 24 hours' storage; whereas four bond systems showed decreased bond strength after 6 months' storage. Bond strengths measured after thermocycling were not predictive of those measured after 6 months' storage.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.