[Investigating the Usefulness of Breath-holding Instructions by Change in the Room Light for the Hearing Impaired].
Аннотация
PURPOSE: In imaging examinations of patients with hearing impairment, poor image quality due to poor respiratory arrest is a problem directly related to diagnostic imaging. Although there have been studies of non-sound instructions, there have been few quantitative studies using changes in brightness. The purpose of this study was to investigate whether changes in room brightness can be used as a respiratory cessation indicator.
METHODS
Fourteen patients were imaged using the 2D FIESTA and LAVA-Flex methods on an MRI system under free respiration, voice-guided cessation of respiration (voice instruction), and room brightness change (light/dark instruction), and compared.
RESULTS
The positional fluctuation of the top of the liver by the 2D FIESTA method showed a large positional fluctuation under free breathing, while the positional fluctuation was small under voice instruction and light/dark instruction. The images obtained by the LAVA-Flex method was significantly degraded by motion artifacts under free breathing, whereas there were fewer motion artifacts under voice instruction and light/dark instruction, and the two were the same with no significant difference in visual evaluation.
CONCLUSION
The visual cessation of respiration instruction by changing the brightness of the room is as effective as the audible cessation of respiration instruction, suggesting that it can be applied clinically.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.