MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.15767

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2022

Outcomes of transvenous lead extraction of very old leads using bidirectional rotational mechanical sheaths: Results of a multicentre study.

Migliore Federico, Pittorru Raimondo, Dall'Aglio Pietro Bernardo, De Lazzari Manuel, Rovaris Giovanni, Piazzi Elena, Dentico Alessia, Ferrieri Alessandra, D'Angelo Giuseppe, Marzi Alessandra, Sawaf Basma El, Bertaglia Emanuele, Iliceto Sabino, Gerosa Gino, Tarzia Vincenzo, Carretta Domenico, Mazzone Patrizio

Многоцентровое исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Lead dwell time >10 years is a recognized predictor for transvenous lead extraction (TLE) failure and complications. Data on the efficacy and safety of TLE using the bidirectional rotational mechanical sheaths in patients with very old leads are lacking. In this multicenter study, we reported the outcomes of transvenous rotational mechanical lead extraction in patients with leads implanted for ≥10 years.

METHODS

A total of 441 leads (median: 159 months [135-197]; range: 120-487) in 189 consecutive patients were removed with the Evolution RL sheaths (Cook Medical, Bloomingtom, IN, USA) and mechanical ancillary tools supporting the procedures.

RESULTS

The main indication for TLE was infection in 74% of cases. Complete procedural success rate, clinical success rate, per lead were 94.8% and 98.2%, respectively. Failure of lead extraction was seen in 1.8% of leads. The additional use of a snare via the femoral approach was required in 9% of patients. Lead dwell time was the only predictor of incomplete led removal (odds ratio: 1.009; 95% confidence interval [CI]: 1.003-1.014; p = .002). Four major complication (2%) were encountered. During a mean time follow-up of 31 ± 27 months, 21 patients (11%) died. No procedure-related mortality occurred. Predictors of mortality included severe left ventricular systolic dysfunction (hazard ratio [HR]: 8.06; 95% CI: 2.99-21.73; p = .001), TLE for infection (HR: 8.0; 95% CI: 1.04-62.5; p = .045), diabetes (HR: 3.7; 95% CI: 1.48-9.5; p = .005), and previous systemic infection (HR: 3.1; 95% CI: 1.17-8.24; p = .022). Incomplete lead removal or failure lead extraction did not impact on survival during follow-up.

CONCLUSION

Our findings demonstrated that the use of bidirectional rotational TLE mechanical sheaths combined with different mechanical tools and femoral approach allows reasonable success and safety in patients with very old leads at experienced specialized centers.

Для цитирования: Migliore Federico, Pittorru Raimondo, Dall'Aglio Pietro Bernardo, De Lazzari Manuel, Rovaris Giovanni, Piazzi Elena, Dentico Alessia, Ferrieri Alessandra, D'Angelo Giuseppe, Marzi Alessandra, Sawaf Basma El, Bertaglia Emanuele, Iliceto Sabino, Gerosa Gino, Tarzia Vincenzo, Carretta Domenico, Mazzone Patrizio. Outcomes of transvenous lead extraction of very old leads using bidirectional rotational mechanical sheaths: Results of a multicentre study.. Journal of cardiovascular electrophysiology. 2022. DOI: 10.1111/jce.15767.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.