Clinical Evaluation of AI100 with Shonit™ versus Manual Microscopy.

Jayaprakash C S, Rao Sandeep, Nair Resmi P, Thomas Blessy Mary, Bhandary P Chandni, Anadure Hema N, Chaitra C

СтатьяМногоцентровое исследованиеСравнительное исследованиеОценочное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

Peripheral blood smear (PBS) is an essential diagnostic test in current medical practice. AI100 with Shonit™ (AI100) is an automatic cell-locating device intended for white blood cell (WBC) differential count, red blood cell (RBC), and platelet morphology precharacterization. We report a multicenter study evaluating the clinical performance of the AI100 with Shonit™ (combined performance after qualified reviewer verification) in PBS analysis, establishing its equivalence to manual microscopy for diagnostic purposes.

MATERIALS AND

METHODS

A retrospective multisite study was conducted at four sites (three in India and one in USA) from August 2021 to January 2023. A total of 882 samples were analyzed by two qualified reviewers in each site independently using manual microscopy and AI100 with Shonit™. The post-review results were compared between AI100 and manual microscopy.

RESULTS

The overall agreement, sensitivity, and specificity for WBC abnormality are 95.03%, 92.68% and 95.41%, respectively. The overall agreement for RBC anisocytosis and poikilocytosis is 94.7% and 92.1%, respectively. For platelet morphology, the overall agreement, sensitivity, and specificity were 96.8%, 97.9%, and 94.6%, respectively. The repeatability and reproducibility of AI100 with Shonit ™ were also evaluated.

CONCLUSION

The performance comparison study results demonstrate that the AI100 with Shonit™ is substantially equivalent to manual microscopy.

Для цитирования: Jayaprakash C S, Rao Sandeep, Nair Resmi P, Thomas Blessy Mary, Bhandary P Chandni, Anadure Hema N, Chaitra C. Clinical Evaluation of AI100 with Shonit™ versus Manual Microscopy.. Indian journal of pathology & microbiology. 2026. DOI: 10.4103/ijpm.ijpm_134_25.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.