Exercise-induced atrioventricular block in a patient with a pacemaker: a functional consequence of AV search algorithms-a case report.

Nuevo-Gallardo Clara, Rojo-Pérez José Miguel, Doncel-Vecino Luis Javier, Muñoz-Santos Lorenzo, Chacón-Piñero Antonio

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Algorithms designed to promote intrinsic atrioventricular (AV) conduction in dual-chamber pacemakers aim to reduce unnecessary right ventricular pacing and its long-term adverse effects. The AV Search+ algorithm periodically prolongs the AV delay to facilitate native conduction and is particularly useful in non-pacemaker-dependent patients. However, under certain conditions, such algorithms may lead to unexpected disturbances in AV synchrony, especially during exercise.

CASE SUMMARY: A 65-year-old woman with symptomatic second-degree AV block (Mobitz II) underwent dual-chamber pacemaker implantation. During follow-up, she developed progressive exertional dyspnoea. Exercise stress testing revealed a paradoxical decrease in heart rate during increasing workload with reproduction of symptoms. Programmer-monitored exercise testing demonstrated a functional 2:1 AV block caused by atrial undersensing, as atrial depolarizations occurred within the post-ventricular atrial refractory period during AV Search+-mediated AV delay prolongation. Deactivation of the AV Search+ algorithm restored stable 1:1 AV conduction during exercise and resulted in complete symptom resolution.

CONCLUSION

This case highlights a rare but clinically relevant complication of AV conduction search algorithms in dual-chamber pacemakers. Although these algorithms effectively reduce ventricular pacing, inappropriate interactions between prolonged AV delays and atrial refractory periods may result in symptomatic functional AV block, particularly during exercise and in patients with advanced AV conduction disease. Exercise testing with device interrogation is a key diagnostic tool in pacemaker patients with exertion-related symptoms, and individualized device programming is essential to ensure optimal clinical outcomes.

Для цитирования: Nuevo-Gallardo Clara, Rojo-Pérez José Miguel, Doncel-Vecino Luis Javier, Muñoz-Santos Lorenzo, Chacón-Piñero Antonio. Exercise-induced atrioventricular block in a patient with a pacemaker: a functional consequence of AV search algorithms-a case report.. European heart journal. Case reports. 2026. DOI: 10.1093/ehjcr/ytag244.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.