Assessing the Accuracy and Reliability of the Fluoroscopic Ruler for Comminuted Femur Fractures: A Cadaveric Study.

Wahlig Brian D, Kuttner Nicolas P, Kouzel-Martinez Frank A, Broida Samuel E, Sems Stephen A, Hidden Krystin A, Yuan Brandon J

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Fixation of comminuted femur fractures may result in limb length discrepancy. Intraoperative fluoroscopic measurement of the contralateral femur with a ruler is commonly performed to establish a reference for femoral length. No evidence regarding the reliability and accuracy of this technique exists. This study aimed to assess the accuracy and interrater reliability of a fluoroscopic ruler in obtaining correct femoral length in a comminuted femoral shaft fracture model.

MATERIALS AND

METHODS

Approximately 5 cm of bone was removed from the left femoral diaphyses of 8 cadavers. Seven orthopedic surgery residents and 2 attendings measured the length of the intact contralateral femur using a ruler under fluoroscopy. The ruler was then applied to the "fractured" femur with manual traction applied until femoral length matched the measured length of the contralateral femur. The resulting gap in the "fractured" femur was compared with the length of bone that had been resected. Data were analyzed using means, SDs, and intraclass correlation coefficients (ICCs).

RESULTS

Fifty-seven measurements were collected. The mean difference between the measured fracture gap and the length of bone removed was 8.0±5.8 mm (range, 0-22 mm). Femoral length was accurate to 5 mm in 40% of cases, 10 mm in 70%, 15 mm in 81%, 20 mm in 98%, and 25 mm in 100%. The overall interrater reliability was poor (ICC, 0.11; 95% CI, 0.001-0.44).

CONCLUSION

Despite poor interrater reliability, the fluoroscopic ruler resulted in a mean leg length discrepancy of 8.0±5.8 mm in this cadaveric study. [Orthopedics. 2024;47(6):327-331.].

Для цитирования: Wahlig Brian D, Kuttner Nicolas P, Kouzel-Martinez Frank A, Broida Samuel E, Sems Stephen A, Hidden Krystin A, Yuan Brandon J. Assessing the Accuracy and Reliability of the Fluoroscopic Ruler for Comminuted Femur Fractures: A Cadaveric Study.. Orthopedics. 2024. DOI: 10.3928/01477447-20240918-03.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.