DOI 10.1097/ccm.0000000000006841
Multimodal Noninvasive Monitoring of Arterial Pressure: Do Limitations of Finger Cuff Monitoring Compromise Upper Arm Measurements? A Post Hoc Analysis of Two Prospective Cohorts.
Аннотация
OBJECTIVES
To assess the reliability of conventional automated oscillometric upper arm cuff (cuff ARM ): 1) when finger cuff monitoring (cuff FINGER ) of arterial pressure (AP) is suspected to be unreliable due to poor fingertip perfusion and/or hand edema or 2) fails to provide a reading.
DESIGN
Prospective observational study based on cohorts evaluating the CNAP (Dräger Medical) and ClearSight (Edwards Lifesciences) systems.
SETTING
Three ICUs.
PATIENTS: Adults with an arterial catheter and stable AP.
INTERVENTIONS: Three sets of triplicate AP measurements-simultaneous readings from the cuff FINGER , cuff ARM and, as a reference, arterial catheter-were collected for each patient.
MEASUREMENTS AND MAIN
RESULTS
We analyzed 352 patients: 214 with the CNAP and 138 with the ClearSight cuff FINGER . In the 102 patients (29%) with poor fingertip perfusion-defined as a capillary refill time greater than 4 seconds-and/or hand edema, cuff ARM mean AP measurements ( n = 304) never failed, met the International Organization for Standardization (ISO) 81060-2:2018 standard (bias ± sd : 2.2 ± 7.2 mm Hg), posed no or low risk of harm in 99.7% of cases, and demonstrated good ability to detect hypotension and hypertension (area under the receiver operating characteristic curve, 0.91 [95% CI, 0.87-0.94] and 0.92 [95% CI, 0.75-1], respectively). In these 102 patients, cuff FINGER failed to display a reading in 17 patients (17%), and if not, did not meet the ISO 81060-2:2018 standard. In 38 of 352 patients (11%) in whom the cuff FINGER failed to display an AP measurement, cuff ARM met the ISO 81060-2:2018 standard for mean AP (bias ± sd : 1.7 ± 7.1 mm Hg). In the absence of failure and risk factors for erroneous measurements, ClearSight met the ISO 81060-2:2018 standard for mean AP, unlike the CNAP system.
CONCLUSIONS
In a multimodal, noninvasive strategy for AP monitoring, ClearSight cuff FINGER may be suitable as a first-line tool due to its continuous nature. When unreliable or unavailable, cuff ARM provides reliable measurements, effectively detects hypotension and hypertension, and poses no significant risk to the patient.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.