Identifying Potential Ventilator Auto-Triggering Among Organ Procurement Organization Referrals.

Henry Nicholas R, Russian Christopher J, Nespral Joseph

СтатьяПеревод в очереди

Аннотация

CONTEXT: Ventilator auto-trigger is the delivery of an assisted mechanical ventilated breath over the set ventilator frequency in the absence of a spontaneous inspiratory effort and can be caused by inappropriate ventilator trigger sensitivity. Ventilator auto-trigger can be misinterpreted as a spontaneous breath and has the potential to delay or prevent brain death testing and confuse health-care professionals and/or patient families.

OBJECTIVE

To determine the frequency of organ donor referrals from 1 Organ Procurement Organization (OPO) that could benefit from an algorithm designed to assist organ recovery coordinators to identify and correct ventilator auto-triggering.

DESIGN

This retrospective analysis evaluated documentation of organ donor referrals from 1 OPO in central Texas during the 2013 calendar year that resulted in the withdrawal of care by the patient's family and the recovery of organs.

MAIN OUTCOME MEASURES: The frequency of referrals that presented with absent brain stem reflexes except for additional respirations over the set ventilator rate was determined to assess for the need of the proposed algorithm.

RESULTS

Documentation of 672 organ procurement organization referrals was evaluated. Documentation from 42 referrals that resulted in the withdrawal of care and 21 referrals that resulted in the recovery of organs were identified with absent brain stem reflexes except for spontaneous respirations on the mechanical ventilator. As a result, an algorithm designed to identify and correct ventilator auto-trigger could have been used 63 times during the 2013 calendar year.

Для цитирования: Henry Nicholas R, Russian Christopher J, Nespral Joseph. Identifying Potential Ventilator Auto-Triggering Among Organ Procurement Organization Referrals.. Progress in transplantation (Aliso Viejo, Calif.). 2016. DOI: 10.1177/1526924816640649.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.