DOI 10.1016/j.apergo.2020.103045
Development of an ergonomic four-finger-push manual pipette design.
Аннотация
The purpose of this study was to develop a new pipette design in which the pipette is operated by four fingers, taking into account the anatomy and anthropometry of the hand. The proposed new pipette designs were compared with a traditional thumb-push pipette based on muscle activity, wrist posture, subjective discomfort ratings for upper extremities, and user preference. The results of the study revealed that the four-finger, ergonomic pipette design reduced muscle exertion (25% reduction for aspiration and 35% reduction for dispensing), awkward wrist posture (33% reduction in wrist flexion, radial and ulnar deviation), and perceived discomfort in the wrist, hand and lower arm. Furthermore, most participants (9 of 10) who used a pipette in their daily work preferred the new pipette designs to a traditional thumb-push pipette design. Thus, we expect that this study will contribute to the reduction of WMSDs risk factors and pain.