Predicting failures of suture anchors used for rotator cuff repair: a CT-based 3-dimensional finite element analysis.
Аннотация
BACKGROUND
Failure of inserted anchors has been recognized as one of the major pathomechanisms of re-tearing after rotator cuff repair.
OBJECTIVE
To predict the inserted anchor failure using CT-based 3-dimensional finite element method (CT/3D-FEM).
METHODS
Among twenty patients who underwent rotator cuff repair, 5 had anchor failure (failed anchor group) and 15 had no anchor failure (stable anchor group). A 3D model of proximal humerus was developed for each patient based on the CT data. A virtual pullout testing of TWINFIX™ anchors inserted into bone at 6 different sites was performed using FEM. Then, mean failure load of 6 anchors for each patient was compared between two groups. Moreover, an optimal cut-off value of the mean failure load was determined for predicting anchor failure.
RESULTS
The mean failure load in the failed anchor group (70.3 N) was significantly lower than that in the stable anchor group (119.0 N; p<0.0001). In our method, the optimum cut-off value of the mean failure load was 75.4 N.
CONCLUSIONS
Failure of the inserted TWINFIX™ anchor could be predicted using CT/3D-FEM. In this method, there seemed to be a high risk of anchor failure in shoulders with a mean failure load of <75.4 N.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.