Condensation in a Membrane Oxygenator Analyzed by Bronchoscopy and Ultrasonography.
Аннотация
OBJECTIVE
To elucidate the mechanism of condensation-related oxygenator dysfunction (CROD) and explore preventive strategies, we directly visualized the interior of a membrane oxygenator using bronchoscopy and ultrasonography and evaluated the effect of external warming.
METHODS
The CROD model consisted of a circuit incorporating a membrane oxygenator perfused with water at 3 L/min and supplied with gas (50% O2 in air) at 2 L/min. The hollow fiber bundle was observed bronchoscopically at the outlet surface and ultrasonographically from the lateral aspect. The percentage droplet area was quantified. In a heating model, the oxygenator was externally warmed with circulating 39°C water. Changes in droplet area, gas-supply pressure, and outlet-surface temperature were analyzed. A single experiment was conducted for each model.
RESULTS
In the CROD model, condensation appeared at approximately 2 h, predominantly in the cooler outer zone. The droplet area increased over time and showed a strong positive association with gas-supply pressure (r = 0.9041). Gas flushing partially reduced both parameters but did not restore baseline values. Ultrasonographic findings remained similar to those of a dry oxygenator. In the heating model, the outer zone was approximately 1°C warmer than the inner zone, and no visible condensation developed over 24 h, with stable gas pressure.
CONCLUSIONS
Two-directional visualization demonstrated that outlet-surface condensation, mainly in the outer zone, was associated with elevated gas-supply pressure in this mechanistic model of CROD. Uniform external heating suppressed condensation, suggesting that temperature control and early detection may represent practical strategies to mitigate CROD.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.