DOI 10.1016/j.ejvs.2020.02.013

Медицинская библиотека · European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery · 2020

Bedside Toe Pressures Measurements in a Department of Vascular Surgery: A Study of Diagnostic Accuracy.

Glargaard Gine L, Høyer Christian, Høgh Annette

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

In patients with peripheral arterial disease (PAD), measurements of distal blood pressure form the basis of assessing the severity of the disease along with symptoms and objective findings. The reliability and accuracy of a fully automated bedside device (SysToe) vs. strain gauge plethysmography (SGP) in patients with low toe pressures in a bedside setting in an outpatient clinic of vascular surgery were investigated.

METHODS

This was a prospective, single blinded study carried out in compliance with the Standards of Reporting of Diagnostic Accuracy Studies. A total of 94 outpatients with symptomatic PAD in an outpatient vascular surgery clinic were included and had index measurements performed with the automated bedside device in a bedside setting by observers with no training in routine distal pressure measurements. SGP reference measurements were performed in a vascular laboratory.

RESULTS

There was agreement between modalities regarding the diagnostic classification of chronic limb threatening ischaemia (CLTI) in 79/94 (84%) patients. For detection of CLTI, Cohen's kappa was 0.57 with a sensitivity of 94%, specificity of 82%, positive predictive value (PPV) 52%, and negative predictive value (NPV) 98%. On average, the automated bedside device underestimated the pressures, compared with the SGP. The mean difference between SGP and SysToe was 9 ± 16 mmHg for right limbs and 10 ± 16 mmHg for left limbs (p < .001).

CONCLUSION

The automated bedside device showed a high sensitivity and a high NPV for excluding CLTI. However, the low PPV may result in overdiagnosis. The automated bedside device may function as a screening tool identifying patients in need of further diagnostics with more accurate equipment, for example SGP.

Для цитирования: Glargaard Gine L, Høyer Christian, Høgh Annette. Bedside Toe Pressures Measurements in a Department of Vascular Surgery: A Study of Diagnostic Accuracy.. European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery. 2020. DOI: 10.1016/j.ejvs.2020.02.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.