Comparison of first deflection annotation algorithm and near-field annotation algorithm in premature ventricular complex ablation.

Pongbangli Natnicha, Ueda Akiko, Nonoguchi Noriko Matsushita, Soejima Kyoko

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Accurate identification of the earliest activation site is essential for successful catheter ablation of premature ventricular complexes (PVCs).

OBJECTIVE

This study aimed to compare first deflection (FD) and near-field (NF) detection using the Turbomap modality on the Ensite X system.

METHODS

A retrospective analysis was conducted on 59 patients (71 ablation sites) who underwent first-time PVC ablation. All cases were remapped using Turbomap to extract FD and NF parameters at the earliest activation sites. Comparisons were based on differences in local activation time (ΔLAT), peak frequency (Δ peak frequency), activation area (Δ activation area), and spatial distance.

RESULTS

FD detection identified significantly earlier LAT (-37.8 ± 14.7 msec vs -27.9 ± 12.4 msec, P < .001). Conversely, NF detection detected a higher peak frequency (249.7 ± 102.8 Hz vs 219.3 ± 103.9 Hz, P = .044). No significant differences were observed in the area of the earliest activation within 5 msec between techniques. When cases were categorized by PVC origin (right ventricular outflow tract vs left ventricular summit) and by ablation strategy (single-chamber vs multi-chamber ablation), no significant differences in mapping parameters were found. However, a shorter distance between the FD- and NF-annotated earliest activation sites was significantly associated with successful PVC elimination within 2 ablation attempts (3.1 ± 2.7 mm vs 8.3 ± 5.0 mm, P < .001).

CONCLUSION

FD detection provides earlier LAT activation, while NF detection captures higher-frequency signals. Closer spatial correlation between FD and NF may predict more efficient ablation, even though mapping parameters did not differ significantly across categorized groups.

Для цитирования: Pongbangli Natnicha, Ueda Akiko, Nonoguchi Noriko Matsushita, Soejima Kyoko. Comparison of first deflection annotation algorithm and near-field annotation algorithm in premature ventricular complex ablation.. Heart rhythm. 2026. DOI: 10.1016/j.hrthm.2026.01.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.