MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11548-026-03654-8

Медицинская библиотека · International journal of computer assisted radiology and surgery · 2026

A novel wearable interface based on IMUs for robotic endoscopy.

Yao Qiwen, Zhang Chengyu, Lin Qiang, Qu Yuesheng, Zhang Chi, Chen Baojun, Zuo Siyang

СтатьяПеревод в очереди

Аннотация

PURPOSE: Flexible endoscopic robots hold broad application prospects in gastrointestinal tract surgery. However, the lack of intuitiveness and heterogeneity between the interface and flexible endoscopic robots results in insufficient operational intuitiveness and prolonged surgery time. Developing an agile, high-precision, and intuitive control method for flexible endoscopic robots has become a challenge.

METHODS

This work proposed a novel method for estimating upper-limb's pose using inertial measurement units (IMUs) by building the upper-limb kinematic model and estimating the pose using solely angle. Additionally, a tremor filter and safety constraints on workspace and velocity were incorporated. Based on this method, this work proposed a novel wearable interface that isomorphically matches to the upper limb and suitable for controlling flexible endoscopic robots.

RESULTS

The wearable interface's positioning root mean square error (RMSE) reaches approximately 1% of full scale in each axis. Under the control of the wearable interface, the planar positioning RMSE of the flexible manipulator reached 0.56 mm. In the ex-vivo experiment, 156 mm 2 of porcine colon mucosa was successfully cut off.

CONCLUSION

The wearable interface has been proven effective in performing endoscopic submucosal dissection (ESD) which is particularly suitable for flexible endoscopic robots. The experimental results demonstrate the robustness, intuitive, and clinical potential of the leader-follower control system.

Для цитирования: Yao Qiwen, Zhang Chengyu, Lin Qiang, Qu Yuesheng, Zhang Chi, Chen Baojun, Zuo Siyang. A novel wearable interface based on IMUs for robotic endoscopy.. International journal of computer assisted radiology and surgery. 2026. DOI: 10.1007/s11548-026-03654-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.