Comparison of surgical techniques for optimal lead placement in sacral neuromodulation: a cadaver study.

Dawoud C, Reissig L, Müller C, Jahl M, Harpain F, Capek B, Weninger W J, Riss S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Sacral neuromodulation (SNM) is a common treatment for patients with urinary and faecal incontinence. A close contact of the tined lead electrode with the targeted nerve is likely to improve functional outcome. The aim of this study was to compare the position of the SNM lead in relation to the sacral nerve by comparing different implantation techniques.

METHODS

This cadaver study was conducted at the Division of Anatomy of Vienna's Medical University in October 2020. We dissected 10 cadavers after bilateral SNM lead implantation (n = 20), using two different standardized implantation techniques. The cadavers were categorized as group A (n = 10), representing the conventional guided implantation group and group B (n = 10), where SNM implantation was conducted with the novel fluoroscopy-guided "H"-technique. The primary goal was to assess the distance between the sacral nerve and the lead placement.

RESULTS

The electrodes were inserted at a median angle of 58.5° (46-65°) in group A and 60° (50-65°) in group B, without reaching statistical significance. In 8 cadavers, the lead entered the S3 foramen successfully. The median distance of the lead to the nerve did not show a significant difference between both groups (E0: Group A: 0.0 mm vs. Group B: 0.0 mm, p = 0.969; E1: Group A: 0.0 mm vs. Group B: 0.5 mm p = 0.754; E2: Group A: 2.5 mm vs. Group B: 2.5 mm p = 1.000; E3: Group A: 3.5 mm vs. Group B: 4.0 mm p = 0.675). In 2 cases (20%) of the conventional group A, the lead was misplaced and located at the gluteal muscle. Perforation of the presacral fascia was observed in one lead placement in group A and in two placements in group B.

CONCLUSIONS

Both standardized implantation techniques may ensure close electrode proximity to the targeted nerve. Misplacement of the electrode was more often observed with the conventional implantation technique.

Для цитирования: Dawoud C, Reissig L, Müller C, Jahl M, Harpain F, Capek B, Weninger W J, Riss S. Comparison of surgical techniques for optimal lead placement in sacral neuromodulation: a cadaver study.. Techniques in coloproctology. 2022. DOI: 10.1007/s10151-022-02632-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.