DOI 10.1583/14-4762.1

Медицинская библиотека · Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2014

Fatigue and in vivo validation of a peritoneum-lined self-expanding nitinol stent-graft.

Bastijanic Jennifer M, Etscheidt Jordan, Sattiraju Mallika, Bonsignore Craig, Kopchok George, White Rodney, Sarac Timur P

СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАВалидационное исследованиеПеревод в очереди

Аннотация

PURPOSE: To assess the fatigue and in vivo performance of a new stent-graft incorporating bovine peritoneum lining that is designed for application in peripheral vascular occlusive disease.

METHODS

Bovine peritoneum-lined stent-grafts were subjected to accelerated in vitro pulsatile fatigue and axial/torsional fatigue testing designed to simulate 10 years of physiological strain on the devices. At specified times the devices were evaluated for stent fracture, suture failure, or tissue tearing. Seven dogs underwent bilateral common iliac artery (CIA) balloon angioplasty injury with unilateral placement of the peritoneum-lined stent-graft. Angiography and intravascular ultrasound were performed prior to treatment, after treatment, and prior to sacrifice at 30 days. Vessels were perfusion fixed and histologically evaluated at 5 regions: above stent, proximal stent, mid stent, distal stent, and below stent.

RESULTS

No evidence of stent, suture, or tissue failure was present during or after pulsatile and axial/torsional fatigue testing. At 30±0.3 days after implantation, all vessels were patent. The average lumen area at explantation across stented vessels was 25.45 mm(2). Lumen areas tended to be reduced above (23.57 mm(2)) and below (24.17 mm(2)) the stent. Lumen areas were consistent across stented regions at explantation (proximal stent 27.80 mm(2), mid stent 25.88 mm(2), and distal stent 25.81 mm(2)). The mean neointimal area in peritoneum-lined stents was 2.02±1.52 mm(2), with a neointima:media ratio of 1.03±0.50. These values were significantly lower in the above and below stent areas than in the stented regions, but there was no difference in either measure within the proximal, mid, or distal stent.

CONCLUSION

The custom-designed peritoneum-lined stent-graft is promising for clinical peripheral applications due to its ability to resist relevant long-term physiological stresses and outstanding short-term patency rates in canine implantations.

Для цитирования: Bastijanic Jennifer M, Etscheidt Jordan, Sattiraju Mallika, Bonsignore Craig, Kopchok George, White Rodney, Sarac Timur P. Fatigue and in vivo validation of a peritoneum-lined self-expanding nitinol stent-graft.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2014. DOI: 10.1583/14-4762.1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.