Lee William G, Evans Lauren L, Chen Caressa S, Fuchs Julie R, Zamora Irving J, Bruzoni Matias, Harrison Michael R, Muensterer Oliver J
Аннотация
INTRODUCTION: Delayed primary repair of esophageal atresia in patients with high-risk physiologic and anatomic comorbidities remains a daunting challenge with an increased risk for peri-operative morbidity and mortality via conventional repair. The Connect-EA device facilitates the endoscopic creation of a secure esophageal anastomosis. This follow-up study reports our long-term outcomes with the novel esophageal magnetic compression anastomosis (EMCA) Connect-EA device for EA repair, as well as lessons learned from the ten first-in-human cases. We propose an algorithm to maximize the advantages of the device for EA repair.
METHODS: Under compassionate use approval, from June 2019 to December 2022, ten patients with prohibitive surgical or medical risk factors underwent attempted EMCA with this device. All patients underwent prior gastrostomy, tracheoesophageal fistula ligation (if necessary), and demonstrated pouch apposition prior to EMCA.
RESULTS: Successful device deployment and EMCA formation were achieved in nine patients (90%). Mean time to anastomosis formation was 8 days (range 5-14) and the device was retrieved endoscopically in five (56%) cases. At median follow-up of 22 months (range 4-45), seven patients (78%) are tolerating oral nutrition. Balloon dilations (median 4, range 1-11) were performed either prophylactically for radiographic asymptomatic anastomotic narrowing (n = 7, 78%) or to treat clinically-significant anastomotic narrowing (n = 2, 22%) with no ongoing dilations at 3-month follow up post-repair.
CONCLUSION: EMCA with the Connect-EA device is a safe and feasible minimally-invasive alterative for EA repair in high-risk surgical patients. Promising post-operative outcomes warrant further Phase I investigation.
LEVEL OF EVIDENCE: IV, Case series of novel operative technique without comparison group.
METHODS: Under compassionate use approval, from June 2019 to December 2022, ten patients with prohibitive surgical or medical risk factors underwent attempted EMCA with this device. All patients underwent prior gastrostomy, tracheoesophageal fistula ligation (if necessary), and demonstrated pouch apposition prior to EMCA.
RESULTS: Successful device deployment and EMCA formation were achieved in nine patients (90%). Mean time to anastomosis formation was 8 days (range 5-14) and the device was retrieved endoscopically in five (56%) cases. At median follow-up of 22 months (range 4-45), seven patients (78%) are tolerating oral nutrition. Balloon dilations (median 4, range 1-11) were performed either prophylactically for radiographic asymptomatic anastomotic narrowing (n = 7, 78%) or to treat clinically-significant anastomotic narrowing (n = 2, 22%) with no ongoing dilations at 3-month follow up post-repair.
CONCLUSION: EMCA with the Connect-EA device is a safe and feasible minimally-invasive alterative for EA repair in high-risk surgical patients. Promising post-operative outcomes warrant further Phase I investigation.
LEVEL OF EVIDENCE: IV, Case series of novel operative technique without comparison group.