Routine application of three-dimensional transesophageal echocardiography-based device selection for transcatheter closure of atrial septal defect without balloon sizing.

Lee Pil Hyung, Kim Ga Yun, Yu Jae Sung, Earm Inbae, Kwon Kyung Beom, Koo Hyung Il, Kim Tae Oh, Lee Sahmin, Sun Byung Joo, Kim Dae-Hee, Kang Duk-Hyun, Song Jae-Kwan, Song Jong-Min

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

AIMS: There is no standardized global framework for device size selection in transcatheter atrial septal defect (ASD) closure. We aimed to evaluate outcomes and insights from the routine application of a device selection strategy based on 3D transesophageal echocardiography (TEE) in consecutive patients undergoing ASD closure.

METHODS AND

RESULTS

The study population was derived from a prospective ASD_3D_Device_Size registry, in which device sizing was exclusively determined prior to the procedure by a 3D TEE-based formula. All patients were clinically followed for a median of 1.6 years after the procedure. Between September 2016 and May 2024, a total of 748 patients were included. Median procedural and fluoroscopy times were 18.0 and 3.7 min, respectively. Procedure success rate was 99.5%. One patient (0.1%) required intraprocedural device size reselection. Device embolization occurred in three (0.4%) patients; two of these cases were attributed to delivery system malfunction. Device malposition was found in one (0.1%) patient, who underwent surgery. During follow-up, one patient required surgery for cardiac erosion. Among 186 (24.9%) oval ASDs, long axis orientations were diverse. Computational simulations demonstrated that, when balloon sizing was assumed to be performed in the fluoroscopic antero-posterior projection, the greatest underestimation of maximal diameter, with a discrepancy of 34.9%, occurred in the horizontally oriented oval ASDs.

CONCLUSION

Pre-procedural determination of device size based on 3D TEE provides a safe and useful framework in routine clinical practice. Because of the diversity in maximal diameter orientation, balloon sizing using a fixed fluoroscopic angle may underestimate the true maximal diameter of an ASD.(ClinicalTrials.gov number NCT02097758).

Для цитирования: Lee Pil Hyung, Kim Ga Yun, Yu Jae Sung, Earm Inbae, Kwon Kyung Beom, Koo Hyung Il, Kim Tae Oh, Lee Sahmin, Sun Byung Joo, Kim Dae-Hee, Kang Duk-Hyun, Song Jae-Kwan, Song Jong-Min. Routine application of three-dimensional transesophageal echocardiography-based device selection for transcatheter closure of atrial septal defect without balloon sizing.. European heart journal. Cardiovascular Imaging. 2026. DOI: 10.1093/ehjci/jeag007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.