Saline vs Contrast for Optical Coherence Tomography During Percutaneous Coronary Intervention: A Prospective Multicenter Study.

Gupta Ankush, Chhikara Sanya, Patrikar Seema, Vijayvergiya Rajesh, Singh Ashutosh Kumar, Newaskar Aditi, Nevali Krishna Prasad, Kasinadhuni Ganesh, Maurya Preetika, Singh Balwinder, Singh Navreet, Bajaj Nitin, Swamy Ajay, Sharma Prafull, Goyal Praveg, Mahesh Nalin K, Akasaka Takashi, Seth Ashok

СтатьяПеревод в очереди

Аннотация

BACKGROUND

OCT improves PCI outcomes compared with angiography alone but requires contrast to displace blood for image acquisition. In patients at risk of contrast-induced nephropathy, minimizing contrast use is clinically important.

OBJECTIVES

The objective of the study was to compare heparinized saline and contrast as flushing media for coronary frequency domain optical coherence tomography (OCT) during percutaneous coronary intervention (PCI).

METHODS

In this prospective, multicenter, observational study, paired contrast and heparinized saline OCT pullbacks were obtained during the same procedure from 166 patients undergoing PCI at 4 tertiary centers in India. A total of 2072 matched pairs of OCT frames (in 259 paired OCT runs) were analyzed by 2 blinded investigators. Vessel dimensions including proximal reference diameter, minimal lumen area, minimum lumen diameter, and distal reference diameter (DRD) were compared. Ethical approval was obtained at all participating institutions, and all patients provided informed consent.

RESULTS

Saline and contrast-based OCT demonstrated no significant differences in proximal reference diameter (2.76 ± 0.59 vs 2.74 ± 0.60 mm; P = 0.65), minimal lumen area (5.33 ± 2.48 vs 5.36 ± 2.52 mm2; P = 0.85), minimum lumen diameter (2.55 ± 0.68 vs 2.56 ± 0.66 mm; P = 0.78), or distal reference diameter (2.54 ± 0.55 vs 2.55 ± 0.55 mm; P = 0.83). Bland-Altman analysis showed good agreement without proportional bias. Interobserver reliability for vessel measurements was moderate to good with Intraclass Correlation Coefficient ranging from 0.69 to 0.79. Sensitivity analysis for clustering confirmed the robustness without proportional bias. No adverse events were observed with saline flushes.

CONCLUSIONS

Heparinized saline provides vessel measurements comparable to contrast media for frequency domain OCT without safety concerns. Saline flushing is a feasible alternative for OCT-guided PCI, particularly in patients at risk of contrast-induced nephropathy.

Для цитирования: Gupta Ankush, Chhikara Sanya, Patrikar Seema, Vijayvergiya Rajesh, Singh Ashutosh Kumar, Newaskar Aditi, Nevali Krishna Prasad, Kasinadhuni Ganesh, Maurya Preetika, Singh Balwinder, Singh Navreet, Bajaj Nitin, Swamy Ajay, Sharma Prafull, Goyal Praveg, Mahesh Nalin K, Akasaka Takashi, Seth Ashok. Saline vs Contrast for Optical Coherence Tomography During Percutaneous Coronary Intervention: A Prospective Multicenter Study.. JACC. Advances. 2026. DOI: 10.1016/j.jacadv.2026.102679.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.