MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.rli.0000167423.27180.54

Медицинская библиотека · Investigative radiology · 2005

Magnetic resonance-guided large-core breast biopsy inside a 1.5-T magnetic resonance scanner using an automatic system: in vitro experiments and preliminary clinical experience in four patients.

Pfleiderer Stefan O R, Marx Christiane, Vagner Jörg, Franke Ralf-Peter, Reichenbach Jürgen R, Kaiser Werner A

СтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVE

The aim of this study was to investigate the feasibility and the precision of magnetic resonance (MR)-guided large-core breast biopsies (LCBB) by using the second prototype of an automatic system (ROBITOM II), which is used to localize lesions while operating at the isocenter of a 1.5-T whole-body scanner.

METHODS AND MATERIALS: In comparison to the first prototype, ROBITOM II is equipped with a dedicated double breast coil and a high-speed trocar setting unit. In vitro experiments (n = 25) with grapefruit phantoms, which contained multiple vitamin E capsules (12 x 7 mm in size) as artificial lesions, were performed. Four patients with MR-detectable breast lesions underwent biopsy. A trocar was positioned in front of the lesion and inserted into the breast. Specimens were harvested with a coaxial technique by using a 14-G core needle biopsy gun.

RESULTS

In all 25 in vitro experiments, capsule material was detected in the specimen cylinder. In 4 patients, the coaxial needle was detected exactly at the expected position. Between 8 and 16 tissue cylinders were harvested. Histologic evaluation resulted in 1 invasive ductal carcinoma and 1 papilloma, which were confirmed after open surgery. One patient who had a proven breast cancer was biopsied for exclusion of multifocal disease. She showed fibrocystic changes, whereas open surgery revealed 3 small areas of ductal carcinoma in situ (DCIS). Another patient showed fibroadenoma after biopsy. This patient is in the follow-up period, which has lasted between 3 and 4 months up until now.

CONCLUSIONS

In this pilot patient study, the feasibility of manipulator-assisted large-core breast biopsy inside a 1.5-T whole-body scanner was demonstrated by using ROBITOM II. The precision of the device was confirmed with in vitro experiments. Although these findings are preliminary and the follow-up period is rather short, they nevertheless represent a successful proof-of-principle of LCBB with ROBITOM II.

Для цитирования: Pfleiderer Stefan O R, Marx Christiane, Vagner Jörg, Franke Ralf-Peter, Reichenbach Jürgen R, Kaiser Werner A. Magnetic resonance-guided large-core breast biopsy inside a 1.5-T magnetic resonance scanner using an automatic system: in vitro experiments and preliminary clinical experience in four patients.. Investigative radiology. 2005. DOI: 10.1097/01.rli.0000167423.27180.54.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.