The Annals of thoracic surgery · 1998
Comparison of tissue reactions in the tracheal mucosa surrounding a bioabsorbable and silicone airway stents.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Korpela A, Aarnio P, Sariola H, Törmälä P, Harjula A
Аннотация
BACKGROUND: Treatment of tracheobronchial stenosis is problematic. Conservative methods include stenting the stenotic area, but an ideal stent has not yet been developed. Bioabsorbable airway stents offer benefits; the extraction of the device is unnecessary, and the airway preserves its normal function after stent resorption. The aim of this study was to examine the suitability of self-reinforced poly-L-lactide as a material for an airway stent.
METHODS: A spiral airway stent made of 0.7-mm wire of self-reinforced poly-L-lactide was implanted operatively in 9 rabbits intratracheally; silicone stents served as controls.
RESULTS: Silicone stents had a tendency to become stenosed with encrustation material and to develop a hyperplastic polyp at both ends. Self-reinforced poly-L-lactide stents were well tolerated and caused no foreign body reaction, and they had a tendency to penetrate into the tracheal wall. They had disappeared at the end of the follow-up of 10 months.
CONCLUSIONS: This experimental study showed that bioabsorbable self-reinforced poly-L-lactide is a promising material for an airway stent for treatment of airway stenosis.
METHODS: A spiral airway stent made of 0.7-mm wire of self-reinforced poly-L-lactide was implanted operatively in 9 rabbits intratracheally; silicone stents served as controls.
RESULTS: Silicone stents had a tendency to become stenosed with encrustation material and to develop a hyperplastic polyp at both ends. Self-reinforced poly-L-lactide stents were well tolerated and caused no foreign body reaction, and they had a tendency to penetrate into the tracheal wall. They had disappeared at the end of the follow-up of 10 months.
CONCLUSIONS: This experimental study showed that bioabsorbable self-reinforced poly-L-lactide is a promising material for an airway stent for treatment of airway stenosis.