Speed of blood withdrawal and accurate measurement of oxygen content in mixed venous blood.

Jaschke Katie, Brown Dianna, Clark Alicia, Doull Sarah, English Ashley, Hoover Nicole, Jones Philip, Klamm David, Odom Chung, Primrose Brenna, Sollars Kristin, Ebberts Marci

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Measurement of mixed venous oxygen saturation helps determine whether cardiac output and oxygen delivery are sufficient for metabolic needs. As recommended by the American Association of Critical-Care Nurses guideline, blood samples for determining mixed venous oxygen saturation are obtained by slowly, in 1 to 2 minutes, withdrawing 1.5 mL of blood from the distal port of the pulmonary artery catheter. In theory, the negative force of rapid withdrawal could pull oxygenated blood from the pulmonary capillary bed, causing falsely elevated saturation values.

OBJECTIVE

To determine if the speed of withdrawal affects oxygen content in blood samples used to measure mixed venous oxygen saturation.

METHODS

The sample consisted of heart failure patients with pulmonary artery catheters admitted to a cardiac intensive care unit. A prospective, randomized, 2 × 2 crossover design was used to compare mixed venous oxygen saturation in blood samples obtained quickly or slowly. A total of 50 sets of saturation values were analyzed. Each set included 1 blood sample obtained slowly, in 1 to 2 minutes, and 1 obtained rapidly, in 5 seconds.

RESULTS

The mean difference in saturation values between the fast and the slow groups was -0.3 (CI, -1.5 to 0.8; P = .55), indicating that no meaningful systematic bias is attributable to fast withdrawal of blood.

CONCLUSIONS

Rapid blood sampling does not falsely elevate measurements of mixed venous oxygen saturation.

Для цитирования: Jaschke Katie, Brown Dianna, Clark Alicia, Doull Sarah, English Ashley, Hoover Nicole, Jones Philip, Klamm David, Odom Chung, Primrose Brenna, Sollars Kristin, Ebberts Marci. Speed of blood withdrawal and accurate measurement of oxygen content in mixed venous blood.. American journal of critical care : an official publication, American Association of Critical-Care Nurses. 2014. DOI: 10.4037/ajcc2014309.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.