Validation Against Polysomnography of a Transthoracic Impedance Sensor for Screening of Sleep Apnea in Heart Failure Patients: A Pooled Analysis of AIRLESS and UPGRADE.

Barbieri Fabian, Adukauskaite Agne, Spitaler Philipp, Senoner Thomas, Pfeifer Bernhard, Neururer Sabrina, Jacon Peggy, Venier Sandrine, Limon Sarah, Ben Messaoud Raoua, Pépin Jean-Louis, Hintringer Florian, Dichtl Wolfgang, Defaye Pascal

СтатьяПеревод в очереди

Аннотация

Background/

Introduction

Cardiac implantable electronic devices and their integrated thoracic impedance sensors have been used to detect sleep apnea for over a decade now. Despite their usage in daily clinical practice, there are only limited data on their diagnostic accuracy.

Methods

AIRLESS and UPGRADE were prospective investigator-driven trials meant to validate the AP scan® (Boston Scientific, Marlborough, MA, USA) in heart failure cohorts. Patients, who either fulfilled the criteria for implantation of an implantable cardioverter-defibrillator (ICD), cardiac resynchronization therapy (CRT), or upgrading to CRT according to most recent guidelines at the time of study conduction, were eligible for enrolment. Sleep apnea and its severity, measured by apnea-hypopnea index (AHI), were assessed by polysomnography. For direct comparison, the apnea sensor-derived AP scan® was used from the identical night.

Results

Overall, 80 patients were analyzed. Median AHI was 21.6 events/h (7.1-34.7), while median AP scan® was 33.0 events/h (26.0-43.0). In the overall cohort, the sensor-derived AP scan® correlated significantly with the AHI (r = 0.61, p < 0.001) with a mean difference (MD) of -12.6 (95% confidence interval (CI) -38.2 to 13.0). Furthermore, the AP scan® was found to correlate well with the AHI in patients with obstructive sleep apnea r = 0.73, p = 0.011, MD -5.2, 95% CI -22.7 to 12.3), but not central sleep apnea (r = 0.28, p = 0.348, MD -10.4, 95% CI -35.4 to 14.6).

Conclusions

In an exclusive heart failure cohort, the AP scan® correlated well with the PSG-derived AHI. A similar correlation was found in most subgroups except for patients suffering from central sleep apnea.

Для цитирования: Barbieri Fabian, Adukauskaite Agne, Spitaler Philipp, Senoner Thomas, Pfeifer Bernhard, Neururer Sabrina, Jacon Peggy, Venier Sandrine, Limon Sarah, Ben Messaoud Raoua, Pépin Jean-Louis, Hintringer Florian, Dichtl Wolfgang, Defaye Pascal. Validation Against Polysomnography of a Transthoracic Impedance Sensor for Screening of Sleep Apnea in Heart Failure Patients: A Pooled Analysis of AIRLESS and UPGRADE.. Journal of clinical medicine. 2024. DOI: 10.3390/jcm13247519.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.