Toric implantable collamer lens for moderate to high myopic astigmatism: 3-year follow-up.
Аннотация
BACKGROUND
To evaluate the 3-year clinical outcomes after toric implantable collamer lens (ICL) implantation for the management of moderate to high myopic astigmatism.
METHODS
Thirty-four eyes of 20 patients who underwent toric ICL implantation were reviewed. All eyes completed 3-year follow-up. Uncorrected (UDVA) and corrected (CDVA) distance LogMAR visual acuities, refraction, endothelial cell density (ECD), and surgical complications were evaluated. Vectorial analysis of astigmatic correction was also done.
RESULTS
A significant improvement in UDVA, CDVA, manifest spherical and cylindrical refraction was observed at 1 week and remained stable after 3 years. Twenty-six eyes (76.5%) gained lines of CDVA, and two eyes (5.9%) showed a loss of 1 line of CDVA. The spherical equivalent (SE) was within ±0.50 D of emmetropia in 18 eyes (52.9%) and within ±1.00 D in 28 eyes (82.4%). Differences between target-induced astigmatism (TIA) and surgically-induced astigmatism (SIA) were statistically significant (p < 0.01), and a trend to undercorrection of the refractive astigmatism was present after 3 years. The magnitude of flattening effect (FE) was found to be significantly lower than the magnitude of TIA (p < 0.01). The magnitude of the torque vector was always positive, with a value below 0.50 D in all cases. No vision-threatening complications were observed during the follow-up.
CONCLUSION
Toric ICL implantation is an effective and safe surgical option that provides a relatively predictable and stable refractive correction of myopic astigmatism. Further improvements are needed to minimize the degree of undercorrection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.