A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices.
Аннотация
BACKGROUND
Various water-based heater-cooler devices (HCDs) have been implicated in nontuberculous mycobacteria outbreaks. Ongoing rigorous surveillance for healthcare-associated M. abscessus (HA-Mab) put in place following a prior institutional outbreak of M. abscessus alerted investigators to a cluster of 3 extrapulmonary M. abscessus infections among patients who had undergone cardiothoracic surgery.
METHODS
Investigators convened a multidisciplinary team and launched a comprehensive investigation to identify potential sources of M. abscessus in the healthcare setting. Adherence to tap water avoidance protocols during patient care and HCD cleaning, disinfection, and maintenance practices were reviewed. Relevant environmental samples were obtained. Patient and environmental M. abscessus isolates were compared using multilocus-sequence typing and pulsed-field gel electrophoresis. Smoke testing was performed to evaluate the potential for aerosol generation and dispersion during HCD use. The entire HCD fleet was replaced to mitigate continued transmission.
RESULTS
Clinical presentations of case patients and epidemiologic data supported intraoperative acquisition. M. abscessus was isolated from HCDs used on patients and molecular comparison with patient isolates demonstrated clonality. Smoke testing simulated aerosolization of M. abscessus from HCDs during device operation. Because the HCD fleet was replaced, no additional extrapulmonary HA-Mab infections due to the unique clone identified in this cluster have been detected.
CONCLUSIONS
Despite adhering to HCD cleaning and disinfection strategies beyond manufacturer instructions for use, HCDs became colonized with and ultimately transmitted M. abscessus to 3 patients. Design modifications to better contain aerosols or filter exhaust during device operation are needed to prevent NTM transmission events from water-based HCDs.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.