MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.28942

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2020

Incidence and fate of device-related left pulmonary artery stenosis and aortic coarctation in small infants undergoing transcatheter patent ductus arteriosus closure.

Tomasulo Catherine E, Gillespie Matthew J, Munson David, Demkin Taylor, O'Byrne Michael L, Dori Yoav, Smith Christopher L, Rome Jonathan J, Glatz Andrew C

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

OBJECTIVES

To evaluate short- and middle-term outcomes after transcatheter patent ductus arteriosus (TC-PDA) closure in small infants, specifically device-related left pulmonary artery (LPA) stenosis and aortic coarctation, risk factors, and changes over time.

BACKGROUND

Recent studies have demonstrated successful transcatheter PDA (TC-PDA) closure in small infants. LPA stenosis and aortic coarctation have been seen after TC-PDA, but it is not clear whether device-related LPA/aortic obstruction persists.

METHODS

A single-center retrospective study of infants ≤4 kg who underwent TC-PDA closure from February 1, 2007 to September 1, 2018 was performed, evaluating the incidence and risk factors for LPA stenosis and coarctation.

RESULTS

Forty-four patients underwent successful TC-PDA with Amplatzer Vascular Plug II (AVPII; n = 30), Amplatzer Duct Occluder II-Additional Sizes (n = 10), Amplatzer Duct Occluder I (n = 3), and coil-filled AVPI (n = 1) devices, all via an antegrade approach. Median birthweight and procedural weight were 890 g (range: 490-3,250) and 2.8 kg (range: 1.2-4.0), respectively. Median follow-up was 0.7 years (range: 2 days-7 years). Thirty-eight patients had post-procedure echocardiograms assessing LPA/aortic obstruction. Of those, 17 had LPA flow acceleration/stenosis (≥1.5 m/s), which improved or resolved in all patients with available follow-up; 3 developed mild coarctation (>2 m/s), which improved in the two with more than short-term follow-up; 4 developed mild flow acceleration (1.5-2 m/s) in the descending aorta, which resolved in three and increased in one (2.4 m/s). Flow acceleration in the LPA was associated with younger procedural age, larger PDA minimal diameter, and placement of a device other than the AVPII. There was no device-related mortality or need for reintervention.

CONCLUSION

TC-PDA in small infants is effective, without significant complications. Device-related LPA/aortic obstruction can improve with time/growth.

Для цитирования: Tomasulo Catherine E, Gillespie Matthew J, Munson David, Demkin Taylor, O'Byrne Michael L, Dori Yoav, Smith Christopher L, Rome Jonathan J, Glatz Andrew C. Incidence and fate of device-related left pulmonary artery stenosis and aortic coarctation in small infants undergoing transcatheter patent ductus arteriosus closure.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2020. DOI: 10.1002/ccd.28942.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.