DOI 10.1016/j.clinph.2019.03.032
SPiQE: An automated analytical tool for detecting and characterising fasciculations in amyotrophic lateral sclerosis.
Аннотация
OBJECTIVES
Fasciculations are a clinical hallmark of amyotrophic lateral sclerosis (ALS). Compared to concentric needle EMG, high-density surface EMG (HDSEMG) is non-invasive and records fasciculation potentials (FPs) from greater muscle volumes over longer durations. To detect and characterise FPs from vast data sets generated by serial HDSEMG, we developed an automated analytical tool.
METHODS
Six ALS patients and two control patients (one with benign fasciculation syndrome and one with multifocal motor neuropathy) underwent 30-minute HDSEMG from biceps and gastrocnemius monthly. In MATLAB we developed a novel, innovative method to identify FPs amidst fluctuating noise levels. One hundred repeats of 5-fold cross validation estimated the model's predictive ability.
RESULTS
By applying this method, we identified 5,318 FPs from 80 minutes of recordings with a sensitivity of 83.6% (+/- 0.2 SEM), specificity of 91.6% (+/- 0.1 SEM) and classification accuracy of 87.9% (+/- 0.1 SEM). An amplitude exclusion threshold (100 μV) removed excessively noisy data without compromising sensitivity. The resulting automated FP counts were not significantly different to the manual counts (p = 0.394).
CONCLUSION
We have devised and internally validated an automated method to accurately identify FPs from HDSEMG, a technique we have named Surface Potential Quantification Engine (SPiQE).
SIGNIFICANCE: Longitudinal quantification of fasciculations in ALS could provide unique insight into motor neuron health.